neuromore Studio

2D Spectrogram

The 2D Spectrogram window shows the frequency content of a device’s EEG channels. Each chart shows the current spectrum, averaged over a short interval. It does not show a history over time. Open it with Windows > Add > 2D Spectrogram. You can open more than one, each with its own settings. It is also part of the Default, Experience Player and EEG layouts.

In 1.7.3: The window has three settings and no display filters, its charts look different, and the Experience Player layout does not include it. See the callouts below and What changed since 1.7.3.

The window reads raw device channels, not classifier output. Filters in your classifier do not change what you see here. The settings in this window do not change the classifier, feedback or other windows.

The charts draw frequencies up to about 60 Hz, or up to half the sample rate if that is lower.

The window’s own display filters are on by default. They can hide real signal, so read the next section before you judge signal quality from this window.

The window has three display filters: High-Pass (Hz), Notch 50 Hz and Notch 60 Hz. They are there because nothing removes electrode drift or mains pickup from a raw channel before it reaches this window.

Each filter sets the frequency bins in its range to zero, with a hard edge:

Filter Bins set to zero
High-Pass (Hz) Below the set frequency
Notch 50 Hz 48 to 52 Hz
Notch 60 Hz 58 to 62 Hz

This has three effects:

  • An empty range does not mean the device sent no signal there. Real signal in these ranges still reaches your classifier.
  • Mains pickup is hidden, so a clean-looking chart does not show that the mains level is low.
  • Auto Scale ignores zeroed bins, so a hidden peak does not change the scale.

The notches do not read NMStudio > Settings > Devices > Local Power Line Frequency. Set them here.

To turn the display filters off:

  1. Click the gear button on the window’s toolbar.
  2. Set High-Pass (Hz) to 0.
  3. Clear Notch 50 Hz and Notch 60 Hz.
  4. If drift or mains peaks flatten the other bands, clear Auto Scale and set Max Amplitude.

In 1.7.3: These filters do not exist. Drift and mains pickup are always shown, and they set the scale.

Click the gear button on the window’s toolbar to open the settings panel. There are nine settings. Defaults can vary by build, so check the values in the panel.

Setting Default Range What it does
Multi View On On: one chart per channel. Off: all channels in one chart, each in its own colour.
Show 0Hz Bin Off Includes the 0 Hz bin. Studio subtracts each FFT window’s mean first, and any high-pass above 0 zeroes this bin.
Average Interval 1.0 0 to 10 Seconds of spectra to average. 0 uses only the latest spectrum.
Horizontal View On Two columns of horizontal bars. Available only when Multi View is on.
Auto Scale On The amplitude scale follows the peak. All charts share one scale.
Max Amplitude 25 1 to 200 Fixes the top of the amplitude scale. Used only when Auto Scale is off. Taller peaks are clipped.
High-Pass (Hz) 0.75 0 to 10 Zeroes bins below this frequency. 0 turns it off.
Notch 50 Hz On Zeroes 48 to 52 Hz.
Notch 60 Hz On Zeroes 58 to 62 Hz.

In 1.7.3: Only Multi View (on), Show 0Hz Bin (off) and Average Interval (1.0, range 0 to 10) exist. The scale always follows the peak, and Studio does not subtract the mean, so the 0 Hz bin carries any signal offset.

The predefined layouts set some of these values. Settings they do not set keep their defaults.

Layout Multi View Show 0Hz Bin Average Interval Horizontal View
Default Off On 1.0 Default
Experience Player On Off 1.0 On
EEG On On 10.0 Default

In 1.7.3: The Experience Player layout does not include the 2D Spectrogram window.

Horizontal View draws each channel’s spectrum as horizontal bars, with frequency on the vertical axis and the lowest frequency at the bottom. Channels fill the left column and then the right column of each row. The two columns are mirrored, so the bars of a pair start from the centre. Bar colour shows the frequency band:

Band Range Colour
Delta Below 4 Hz Red
Theta 4 to 8 Hz Orange
Alpha 8 to 13 Hz Yellow
SMR 13 to 15 Hz Turquoise
Beta 15 to 30 Hz Blue
Gamma 30 Hz and above Purple

With Horizontal View off, Multi View arranges vertical bars in a grid, in each channel’s colour.

In 1.7.3: Horizontal View does not exist. Multi View always uses the grid.

With Multi View and Horizontal View both on, the code draws charts in pairs. If an odd number of channels is selected, the last channel gets no chart. With one channel selected, including on a one-channel device, no chart is drawn. Both settings are on by default, and in the Experience Player and EEG layouts.

To see every selected channel, do one of these:

  • Clear Horizontal View. All channels appear in the grid.
  • Clear Multi View. All channels appear in one chart.
  • Select an even number of channels.

In 1.7.3: Every selected channel is drawn.

The toolbar has a device list, with Use Active Device as the first entry, and a checkbox for each EEG channel of that device. By default the first eight channels are checked. REF and GND are not.

If the active classifier has a Select Channels node with Quick Config on, that node sets these checkboxes on every update, so changes you make here do not stay. Change the selection in the node instead. See Nodes.

In 1.7.3: The checkboxes do not follow the classifier. You choose channels here.

The window takes its FFT settings from the Spectrum Analyzer Settings window: FFT Order, Shift and Window Function. The defaults are order 9 (a 512-point FFT), a shift of 8 samples and a Hann window. These settings apply to every 2D Spectrogram window. They don’t change FFT nodes in a classifier, because each FFT node has its own settings. See Windows, menus and shortcuts.

In 1.7.3: The default FFT order is 7 (a 128-point FFT), so each bar covers a wider frequency range. The default window is Rectangular.